The LifeStream™ Balloon Expandable Vascular Covered Stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries with reference vessel diameters between 4.5 mm and 12.0 mm, and lesion lengths up to 100 mm.
The LifeStream™ Balloon Expandable Vascular Covered Stent was developed using BD's vast experience in PTA and covered stents to create a device designed for the challenging anatomy of iliac arteries and engineered to facilitate accurate placement.
** 5 mm - 8 mm diameters utilize a 37 mm base stent platform. 9 mm - 12 mm diameters utilize a 38 mm base stent platform.
1Ability to post-dilate beyond nominal diameter to help conform to varying patient vessel diameters. 5-8 mm devices can be post-dilated to balloons up to 10 mm in diameter and 9-12 mm devices can be post-dilated with balloon s up to 12 mm in diameter.
2 Foreshortening is calculated as the difference, represented as a percentage, between the labeled covered stent length in crimped condition and the actual stent length measured at both nominal and at rated burst pressure. Across all balloon sizes, foreshortening ranged from -1.5%% to 11.6% at nominal pressure, with a mean of 3.5%, and from -0.8% to 11.8% at rated burst pressure, with a mean of 4.6%. Please consult package insert for the LifeStream™ Covered Stent foreshortening chart.
3 Acute Technical Success defined as successful deployment of the LifeStream™ Covered Stent at the intended location, as determined by the investigator. BOLSTER Clinical Study. Data on File. Bard Peripheral Vascular, Inc., Tempe, AZ.