Indications For Use

The PowerPort™ implantable port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

When used with a PowerLoc™ Safety Infusion Set, the PowerPort™ implantable port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.

 

Contraindications

This device is contraindicated for:

• Catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates of pinchoff. Port may be placed in lateral subclavian vein based on evaluation by a qualified practitioner.

• When the presence of device-related infection, bacteremia, or septicemia is known or suspected.

• When the patient’s body size is insufficient for the size of the implanted device.

• When the patient is known or is suspected to be allergic to materials contained in the device. The device is primarily comprised of silicone, polyacetal, polyetheretherketone, and/or titanium.

• If severe chronic obstructive lung disease exists.

• If the prospective insertion site has been previously irradiated.

• If the prospective placement site has previously suffered episodes of venous thrombosis or vascular surgical procedures.

• If local tissue factors will prevent proper device stabilization and/or access.

 

Warnings

I. During Placement:

• Intended for Single Use. Do not reuse. Reuse and/or repackaging may create risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure and/or lead to injury, illness or death of the patient.

• Alcohol should not be used to soak or declot a polyurethane catheter because alcohol is known to degrade the polyurethane catheter over time with repeated and prolonged exposure.

• After use, this product may be a potential biohazard. Handle and discard in accordance with accepted medical practice and applicable local, state and federal laws and regulations.

• Place thumb over exposed opening of sheath or needle or attach syringe filled with sterile normal saline solution to minimize blood loss and prevent air embolism. The risk of air embolism is reduced by performing this part of the procedure with the patient performing the Valsalva maneuver and/or in Trendelenburg position.

• Do not suture catheter to port, port stem, or surrounding tissue. Any damage or constriction of catheter may compromise power injection performance and catheter integrity. Bard Access Systems, Inc. does not recommend suturing around the catheter as doing so could compress, kink, or damage catheter, including catheter fragmenting and/or fracturing.

• Do not manipulate a pre-assembled or pre-connected catheter/port connection, as the catheter could become disconnected from the port, or system damage could occur.

• Do not attempt to measure the patient’s blood pressure on the arm in which a peripheral system is located, since catheter occlusion or other damage to the catheter could occur.

• Avoid vessel perforation.

• Do not power inject through a port system that exhibits signs of clavicle-first rib compression or pinch-off as it may result in port system failure.

• For implantable ports with Groshong® catheters, do not cut stylet. Withdraw stiffening stylet from catheter prior to cutting.

• Failure to completely advance the catheter on the dual lumen stem may result in subcutaneous leakage.

II. During Port Access:

• Do not use a syringe smaller than 10 mL. Flushing occluded catheters with small syringes can create excessive pressures within the port system.

• PowerPort™ implantable ports are only power injectable when accessed with a PowerLoc™ Safety Infusion Set.

• Failure to warm contrast media to body temperature prior to power injection may result in port system failure.

• Failure to ensure patency of the catheter prior to power injection studies may result in port system failure.

• Power injector machine pressure limiting feature may not prevent over pressurization of an occluded catheter.

• Exceeding the maximum flow rate may result in port system failure and/or catheter tip displacement.

• PowerPort™ implantable port indication for power injection of contrast media implies the port’s ability to withstand the procedure, but it does not imply appropriateness of the procedure for a particular patient nor for a particular infusion set. A suitably trained clinician is responsible for evaluating the health status of a patient as it pertains to a power injection procedure and for evaluating the suitability of any infusion set used to access the port.

• Do not exceed a 300 psi pressure limit setting on the power injection machine, or the maximum recommended flow rate on the PowerLoc™ needle, if power injecting through the PowerPort® implantable port.

• If local pain, swelling or signs of extravasation are noted during power injection, the injection should be stopped immediately.

 

Possible Complications

The use of a subcutaneous port provides an important means of venous access for critically ill patients. However, the potential exists for serious complications, including the following:

Air Embolism, Allergic Reaction, Bleeding, Brachial Plexus Injury, Cardiac Arrhythmia, Cardiac Puncture, Cardiac Tamponade, Catheter or Port Erosion Through the Skin, Catheter Embolism, Catheter Occlusion, Catheter or port-related Sepsis, Damage or Breakage due to Compression between the Clavicle and First Rib, Device Rotation or Extrusion, Endocarditis, Extravasation, Fibrin Sheath Formation, Guidewire Fragment Embolism, Hematoma, Hemothorax, Hydrothorax, Infection, including but not limited to, pocket, catheter tunnel, and/or blood stream, Inflammation, Necrosis, or Scarring of Skin Over Implant Area, Intolerance or Reaction to Implanted Device, Laceration of Vessels or Viscus, Pain at or around port pocket site, Perforation of Vessels or Viscus, Pneumothorax, Risks Normally Associated with Local or General Anesthesia, Surgery, and Post-Operative Recovery, Spontaneous Catheter Tip Malposition or Retraction, Thoracic Duct Injury, Thromboembolism, Vascular Thrombosis, Vessel Erosion

Please consult product labels and package inserts for indications, contraindications, hazards, warnings, cautions, possible complications, and information for use.

 

BD-90736

PowerPort™ ClearVUE™ Implantable Ports

The port designed with your patient's journey in mind.

ClearVUE graphic
Loading

Developed to Redefine
Your Port Expectations

Whether minimizing radiation deflection or avoiding imaging artifact, the ingenious PowerPort™ ClearVUE™ Implantable Port was developed to not only provide you performance at the highest levels, but also to work in conjunction with your patient’s treatment journey. The technologically advanced ClearVUE™ Ports are the first and only metal-free, MR safe, power-injectable port on the market.*

    Clear Imaging

    The metal-free construction of ClearVUE™ Ports enables uninterrupted imaging, minimizing related artifact that otherwise may mask the identification of tumor growth, new cancerous lesions, or metastasized cancer during MRI, CT, and other radiographic imaging.

    Port Artifact Under MRI Scan (3 Tesla)

     

    • ClearVUE graphic

      PowerPort™ ClearVUE™ isp Device

    • ClearVUE graphic

      Plastic PowerPort™ isp Device

    • ClearVUE graphic

      Titanium PowerPort™ isp Device

    Port Artifact Under CT Scan

     

    • ClearVUE graphic

      PowerPort™ ClearVUE™ isp Device

    • ClearVUE graphic

      Plastic PowerPort™ isp Device

    • ClearVUE graphic

      Titanium PowerPort™ isp Device

     

     

    Learn More

     

    BD PowerPort™ ClearVUE™ Slim and isp Implantable Ports were characterized for CT and MRI scan visibility against competitors.

     

    Download Our Brochure

    English
    English
    ClearVUE graphic

    Targeted Radiation

    PowerPort™ ClearVUE™ Ports are engineered uniquely to help you avoid reduction in overall radiation dosage through the port, minimize inadvertent healthy tissue radiation, and ultimately eliminate the need for port relocation to effectively deliver radiation therapy.1

    In the presence of metal, radiation scatters. The metal-free design of PowerPort™ ClearVUE™ Port has demonstrated no observable back scatter, side scatter, or attenuation.2

    ClearVUE™ For Port Placers, Nurses, & Patients

    ClearVUE™ Ports give you distinct advantages in placement, enable uninterrupted imaging and treatment, and provide benefits specifically for both your care team and the patients you care for.

     

    powerport-clearvue-implantable-ports

    Simplified placement & ongoing benefits

     

    • Ergonomic triangular shape is designed for ease of insertion
    • Ipsilateral placement to the treatment area
    • Metal-free construction makes it MR Safe & CT compatible and minimizes imaging artifact as well as radiation therapy interference
    • Durable plastic port and ChronoFlex™ Polyurethane Catheter

     

    Efficient access & administration for your nursing team

    Comfort & convenience for your patients

    References

    * As of June 2026

    1 Compared to metal ports.

    2 Bench test data on file. BD, Tempe, AZ. Bench test data may not correlate to actual clinical outcomes. Different test methods may yield different results.

    ©2026 Illustrations by Mike Austin

    Please consult product labels and inserts for indications, contraindications, hazards, warnings, precautions and directions for use.

    BD-90740

    Clinician Experience

    Engineered with Port Placers in Mind 

    Simplified Placement

    • Ergonomic triangular shape designed to facilitate port pocket insertion
    • Enables port placement ipsilateral to the treatment area
    • Supports right-sided port placement, which is known to have fewer complications compared to placing ports on the left

     

    Ongoing Benefits

    • Metal-free construction
    • MR Safe and CT compatible
    • Clear CT markers
    • Reduced artifact for clear MRI and CT imaging1
    • Minimal interference with radiation therapy1
    • Durable plastic port and ChronoFlex™ Catheter
    • Power-injectable at 5 mL/s at 300 psi

    Designed with Nurses in Mind

     

    Efficient Access & Administration

    • Bench tested to 208 infusions and 36 power injections2
    • Palpation bumps and triangular shape indicate power-injection and assist with septum identification
    • Compatible with infusion of medications including anti-cancer medicines (chemotherapy), contrast media, IV fluids, parenteral nutrition solutions, and blood products
    • Our updated instructions for use now support saline flushing and locking—making routine maintenance safer, easier, and just as effective     

     

    Training & Maintenance Resources

    • Training in-services available nationwide
    • PortReady™ educational material for both providers and patients
    • Visit PortReady.com to learn more

     

     

    Download Our Brochure

    English
    English

    ChronoFlex™ Catheter

    In bench top testing, polycarbonate-based polyurethanes, such as Chronoflex™, exhibited significantly higher tensile strength and burst strength when compared to polyether-based urethanes like Tecoflex™3.

     

    Scanning Electron microscope images of 100X magnification at 6 weeks of implantation with no exposure to Taxol or Taxotere chemotherapy.

    Port Specifications

    All kits include StruXure™ Guidewire 0.035" and AirGuard™ Valved Introducer

     

    References

    1 Compared to metal ports.  Bench test data on file. BD, Tempe, AZ. Test results may not correlate to actual clinical performance. Different test methods may yield different results.

    2 Bench test data on file. BD, Tempe, AZ. Test results may not correlate to actual clinical performance. Different test methods may yield different results.

    3 15 samples each of tubing representing both polyether (Tecoflex) and polycarbonate (Carbothane) urethanes of the same durometer and size were soaked in saline and Taxol and Taxotere carrier solutions for 15 minutes prior to testing. The tensile testing, which measured axial stress by stretching the sample longitudinally, showed that the polyether group (EG) had an overall average 36% reduction in tensile strength compared to the polycarbonate group (PC). The burst strength testing, which measured the hoop or radial strength of the sample, showed that the Polyether group (EG) had an overall average 23.8% reduction in burst strength compared to the polycarbonate group (PC). Bench test data on file, BD, Tempe, AZ. Bench test results may not be indicative of actual clinical performance. Different test methods may yield different results.

    ©2026 Illustrations by Mike Austin

    Please consult product labels and inserts for indications, contraindications, hazards, warnings, precautions and directions for use.

    BD-90740

    Patient Experience

    Created with Patients in Mind

    Comfort & Convenience

    PowerPort™ ClearVUE™ Implantable Ports are available in two sizes to enhance patient comfort. The PowerPort™ ClearVUE™ isp is an intermediate sized port, while the PowerPort™ ClearVUE™ Slim is a low-profile port.

     

    ClearVUE graphic
    • Silicone overmolded port body for a softer feel1
    • Lightweight design
    • Power-injectable
    ClearVUE graphic
    • Low-profile size allows for small incisions and port pockets
    • Ideal for thin or cosmetically-minded patients
    • Lightweight design
    • Power-injectable
    ClearVUE graphic

    PortReady Resources

    • PortReady™ educational material for both providers and patients
    • Visit PortReady.com to learn more or download our brochure below

    Download Brochure

    English
    English

    References

    1 Compared to titanium port body.

    ©2023 Illustrations by Mike Austin

    Please consult product labels and inserts for indications, contraindications, hazards, warnings, precautions and directions for use.

    BD-90740

     

    Created with Port Placers, Nurses and Patients in Mind

    What Other Placers
    Are Saying


    "My staff and I have used a wide range of ports in the past. Once we placed the PowerPort™ ClearVUE™ Implantable Port and saw the metal free and power injectable features firsthand, it was a simple choice to continue using this port in my practice."

    "If I have a patient who is eligible for proton beam therapy to structures in the chest, I will only place a 100% metal free port so that therapy can be delivered without relocating the port. Since the PowerPort™ ClearVUE™ Implantable Port is MR Safe, this allows my staff and I to place the device without concern about interference with future MR imaging."

    -Ripal Gandhi, MD, FSVM, FSIR

     

     

     

    "What truly sets the PowerPort™ ClearVUE™ Implantable Port apart is its intentional design that considers the patient's future diagnostic and treatment needs. The device's 100% MRI safety enables seamless integration with cutting-edge MR therapies, including MR Linac, expanding treatment options and enhancing the quality of care we can provide to our patients."

    -Jon C. Davidson MD, FSIR

    What Other Nurses
    Are Saying

    "What I always do is check if it's a PowerPort™. If it is, I take the opportunity to teach new nurses about how simple the port is to access. You just identify the three bumps, hold it and access right in the middle of the bumps every single time. I do think that is one aspect that makes it a little easier to train new nurses who may not be as comfortable touching or accessing a port."

    "Patients who require frequent access may want to consider an implantable access device of some type. I think it's great to avoid the discomfort associated with repeated PIV access."

    -Keilye Pampillonia, BSN, OCN

    You can purchase the PowerPort™ ClearVUE™ Ports through your local sales representative.

    Contact yours today!

     

    Contact Us

    Ready to Order?

    References

    Please consult product labels and inserts for indications, contraindications, hazards, warnings, precautions and directions for use.

    BD-90740

    Products & Accessories
    Sort by:
    References

    ©2023 Illustrations by Mike Austin

    Resources
    Card Image

    Our collection of literature on industries and our product offerings give you information you can use to continue striving for excellence.

    Resources
    Card Image

    Our eIFU portal allows you to search for a particular product's eIFU. Consult with your representative or check the product labeled IFU if no eIFU is found.

    eifu.bd.com
    true
    PowerPort™ ClearVUE™ Implantable Ports BD Implantable port devices PowerPort™ ClearVUE™ Implantable Ports PowerPort™ /content/dam/bd-assets/bd-com/en-us/logos/bd/header-bd-logo.svg

    The port designed with your patient's journey in mind.

    1678301 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1678300 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1676301 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1676300 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1668362 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details 1666362 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details 1618070 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1618001 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1618000 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1616071 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1616070 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1616001 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1616000 PowerPort™ ClearVUE™ Slim Implantable Port See Product Basic Specification for details 1608062 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details 1608052 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details 1606062 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details 1606052 PowerPort™ ClearVUE™ isp Implantable Port See Product Basic Specification for details
    Power ports, cancer, power injectable, single lumen, heparin locks, MRI Safe Ports, non-metallic ports, implantable port, lightweight port